Contacting Conductivity Sensor for Low Conductivity, K=0.5, 3/4" Kynar Compression FittingUltimate accuracy from ultra-pure to high conductivity applications.Nominal Cell Constant (K) = 0.5, 3/4 inch NPT Kynar (PVDF) Compression Fitting, 6m (20ft) Analog Cable, Titanium ElectrodeHigh Performance Des
Dapatkan penawaran terbaik untuk Contacting Conductivity Sensor for Low Conductivity K=0 5 3/4 Kynar Compression Fitting !
Name
Quantity
Unit
We are Online Media consists of Distributor, Reseller, Manufacture, Supplier. If you are interested to buy Contacting Conductivity Sensor for Low Conductivity K=0 5 3/4 Kynar Compression Fitting cheap, please contact the Company directly.

If you are a Company who wish to sell your product,
Search Contacting Conductivity Sensor for Low Conductivity K=0 5 3/4 Kynar Compression Fitting
Product Details

Contacting Conductivity Sensor for Low Conductivity, K=0.5, 3/4" Kynar Compression Fitting

Ultimate accuracy from ultra-pure to high conductivity applications.

Nominal Cell Constant (K) = 0.5, 3/4 inch NPT Kynar (PVDF) Compression Fitting, 6m (20ft) Analog Cable, Titanium Electrode

  • High Performance Design
  • Resistivity and Conductivity Measurement Capability
  • Versatile Mounting Styles
  • Full-Featured “Plug and Play” Hach sc Digital Controller

Parameter/Range/Reagent Information

ConductivityMethodRangeReagent Set(s)USEPA Compliant?
 Process Conductivity1 - 1000 µS/cm 

Alat Laboratorium Umum

High Performance Design

These sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Also, each sensor has a PT1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1 ºC accuracy.

Resistivity and Conductivity Measurement Capability

These enhanced performance sensors measure from theoretically pure water (0.057 µS/cm or 18.2 MΩ) up to 200,000 µS/cm. Hach’s sc Digital Controllers accept multiple digital sensor inputs, and can be user-set to measure conductivity, resistivity,TDS, salinity, or one of six
calculated measurements.

Versatile Mounting Styles

Compression fitting sensors—Feature titanium electrodes and a compression fitting for universal installation with up to 4 inches (102 mm) insertion depth. The 1/2-inch or 3/4-inch male NPT compression fittings are offered in Kynar® (PVDF) or 316 stainless steel. A longer version of this sensor is available for use with a 316 stainless steel ball valve hardware assembly to insert/retract the sensor from the process without stopping the flow. The longer version can also be used for insertion through a compression fitting. Maximum insertion depth is 7 inches (178 mm).

Full-Featured “Plug and Play” Hach sc Digital Controllers

There are no complicated wiring or set up procedures with any Hach sc controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”

 What's in the box?

Includes: sensor with 20 ft cable and manual

 Specifications

Accuracy:±2 % of reading above 200 µS/cm
Cable Length:20 fixed cable (6 wire, 4 conductors and two isolated shield wires)
Cell Constant k:0.5
Diameter:12.7 mm
Digital Gateway:None
Electrode Material:Titanium
Flow:0 - 3 m/s maximum, fully immersed
Length:193.5 mm
Material:Compression Fitting: KYNAR (PVDF)
Measurement Range 2:0.001 - 20 MΩ
Operating Temperature Range:-20 - 200 °C
Pressure Range:1.7 bar at 150 °C
Range:0 - 1000 µS/cm
Region:Global
Repeatability:±0.5 % of reading
Response Time T90:30 s
Sensitivity:±0.5 % of reading
Sensor Immersion Depth:155 mm
Sensor Thread:½" NPT
Sensor Type:Analog
Temperature Measurement:-20 - 200 °C
Temperature Sensor:PT1000 RTD
Transmission Distance:100 m maximum;
1000 m maximum, when used with a termination box